Measuring of pressure in air and non-corrosive gases. Transduction in voltage and current values.
The sensor is ideal for registering low air flow in air conditioning systems and for the measurement of fine pressures in environmental, laboratory and clean-room applications.
Features
suitability for static and dynamic measurements
short circuit-proof and protected against polarity reversal
fast and easy mounting.
Spec.
sensing type:ceramic fulcrum lever element
Output:4-20mA
Power supply:15-35Vdc
Accuracy:+/-0.7%FS
Dimensions:92mm x 85mm x 49.1mm
Ambient temperature:0ºC - +50ºC
Medium temperature:0ºC - +70ºC
Dynamic:response time < 10oms
Dynamic:pressure cycle < 10 Hz
Weight:130g
Protection:IP54
Class:I
Order Code
Order Code |
Range |
Overload Pressure |
PA21 |
0-100Pa |
5000Pa |
PA22 |
0-300Pa |
5000Pa |
PA23 |
0-500Pa |
10000Pa |
PA24 |
0-1000Pa |
10000Pa |
PA25 |
0-1600Pa |
10000Pa |
PA26 |
0-2500Pa |
20000Pa |
PA27 |
0-5000Pa |
20000Pa |